Vernier LabPro User’s Manual
A-1
Appendix A: Troubleshooting Tips
Refer to http://www.vernier.com/til/1613/ for additional troubleshooting tips.
Problem Cause/Solution
When LabPro is used with the
computer the Logger Pro 3
Collect button is faded (or the
Logger Pro 2 Collect button
does not appear).
The Collect button does not appear because the computer
could not communicate with the interface. Check each of the
following:
Make sure the cables are firmly attached.
Make sure LabPro is connected to the correct port on the
computer.
Confirm that the port on the computer is working properly.
Make sure the interface has power—either AC or battery.
When LabPro is used with a
TI graphing calculator, I get a
link error message
This message appears when the calculator cannot communicate
with the interface. Check the following:
Make sure the cables are firmly attached to the calculator
and LabPro.
Make sure the interface has power—either AC or battery.
The sensor is not working. This may be happening because the sensor is not plugged into
the correct port. Since LabPro has 4 analog and 2 digital ports,
make sure that the sensor is connected to the proper port.
The readings are noisy. LabPro may be picking up interference from other electrical
devices, such as a computer. Move the LabPro interface at
least 30 cm from another computer.
When LabPro is plugged into
AC power, the lights and
beeps are not generated.
When LabPro powers up, lights and beeps signal a successful
self-test. If batteries are already installed in LabPro, the self-
test will not be performed.
When connected to USB, a
light can be seen blinking
inside the interface.
This is normal when LabPro is used through a USB port.